(713) 943-1760
3208 Federal Rd., Pasadena, TX 77504
(713) 943-1760 · TheXrayCompany.com

Liquid Penetrant Testing (PT) Study Guide

Short lessons covering what a PT Level II candidate needs to know, from capillary action to acceptance criteria. Concepts follow NAS 410, SNT-TC-1A, ASTM E1417 / E165, ASME V Article 6, and AWS D17.1.

Take a practice quizAll methods

1. How penetrant testing works

Liquid penetrant testing (PT) finds discontinuities that are open to the surface of nonporous materials: metals, many ceramics and plastics, and glass. A liquid with low surface tension and good wetting is drawn into openings by capillary action. After the surface excess is removed, a developer draws the trapped penetrant back out so it can be seen.

The basic process

  1. Pre-clean and dry the surface
  2. Apply penetrant and allow the dwell time
  3. Remove excess surface penetrant
  4. Dry (when required by the method and developer)
  5. Apply developer and allow developing time
  6. Inspect under the required lighting and evaluate
  7. Record results and post-clean

Properties that matter

  • Wetting and surface tension decide whether the penetrant enters an opening. A low contact angle means good wetting.
  • Viscosity mainly affects how fast it penetrates and drains, not whether it can.
  • Specific gravity is usually below 1, so water contamination tends to settle in open tanks.
  • Fluorescence: Type I dyes absorb UV-A (about 365 nm) and emit yellow-green light the eye sees easily.
  • Flash point is a fire-safety property.
PT cannot find subsurface discontinuities and cannot measure depth. Porous materials give heavy background and are poor candidates.

Back to lessons

2. Penetrant types and removal methods

Penetrant systems are classified (ASTM E1417 and E165/SE-165, using AMS 2644 designations) by type, method, sensitivity level, developer form, and solvent remover class.

Penetrant types
TypeDescriptionViewing
Type IFluorescentUV-A light in a darkened area
Type IIVisible (color contrast), usually redWhite light, against white developer
Removal methods
MethodRemovalStrengthsWatch out for
AWater-washable (emulsifier built in)Fast, economical, good on rough surfaces and production workOver-washing shallow, wide discontinuities
BPost-emulsifiable, lipophilic (oil-based emulsifier, used as supplied)Controlled removalEmulsifier contact time is critical; do not agitate or brush
CSolvent-removablePortable, good for spot checks and field workLabor intensive; never flush with solvent
DPost-emulsifiable, hydrophilic (water-based remover)Controlled removal; high sensitivity on smooth surfacesRemover concentration and contact time must be controlled

Solvent removers: Class 1 halogenated, Class 2 nonhalogenated, Class 3 specific application.

Penetrant, emulsifier or remover, and developer are qualified together as a family. Do not mix families or manufacturers.

Back to lessons

3. Sensitivity levels

Type I penetrants are assigned sensitivity levels. Type II visible penetrants do not have a level.

LevelDescription
1/2Ultralow
1Low
2Medium
3High
4Ultrahigh
GCT practice: GCT Inspection uses Level 3 (high) sensitivity penetrant unless the customer's specification or procedure requires otherwise.

Higher sensitivity is not always better. On rough or porous-looking surfaces, a very sensitive penetrant can leave background that hides indications. Choose the level the specification and procedure call for.

Back to lessons

4. Pre-cleaning and surface preparation

Most PT failures start here. A discontinuity filled with oil, water, scale, paint, or old penetrant cannot take in new penetrant.

  • Oils and greases: vapor degreasing, solvent cleaning, or alkaline cleaners.
  • Scale, rust, and corrosion: chemical descaling or pickling, or a mechanical method the procedure allows.
  • Paint and coatings: strip completely, rinse, and dry.
  • Ultrasonic cleaning helps remove soils from crevices.
  • Dry completely after any aqueous or solvent cleaning. Liquid left in a discontinuity blocks penetrant.

Smeared metal

Grit blasting, shot peening, sanding, grinding, machining, and aggressive power wire brushing can smear metal over openings, especially on soft alloys like aluminum. Many specifications require etching after these operations, followed by thorough rinsing or neutralizing and drying.

As-welded, as-cast, as-rolled, and as-forged surfaces are often acceptable, but grinding or machining may be needed where surface irregularities could mask indications. Under ASME V, the examination area plus an adjacent band must be clean and dry.

Back to lessons

5. Applying penetrant and dwell time

Penetrant may be applied by dipping, brushing, or spraying. The surface must stay wet for the dwell time.

  • Temperature: codes set a standard temperature range for the part and penetrant. Cold makes penetrant slow; heat can dry it and fade fluorescent dye. Processing outside the range needs a qualified procedure.
  • Dwell time depends on the specification, material, product form, expected discontinuity, and temperature. Use the dwell in the governing procedure.
  • Do not let penetrant dry on the surface. If a maximum dwell is exceeded, follow the procedure (often reapply or reprocess).

The specific temperatures and times used by common codes are in the code values table below. Always follow the governing document and the edition your procedure references.

Back to lessons

6. Removing excess penetrant and drying

  • Method A: rinse with a coarse water spray. Check background under UV-A for fluorescent penetrants. Avoid over-washing.
  • Method B: apply lipophilic emulsifier by dip or flow, without agitation. Hold for the specified time, then rinse.
  • Method D: water prerinse to remove the bulk of the penetrant, hydrophilic remover at the specified concentration and contact time, then a final rinse.
  • Method C: wipe off with a clean dry lint-free cloth, then lightly wipe with a cloth dampened (not soaked) with remover. Never spray remover onto the part to flush it.

Rinse water pressure and temperature are limited because harsh rinsing pulls penetrant out of discontinuities. Drying temperature and time are limited so penetrant is not baked out. Apply developer as soon as the procedure allows after drying.

Back to lessons

7. Developers

Developer acts like a blotter. It draws penetrant out, spreads it so the indication is larger than the discontinuity opening, and gives a contrasting background.

FormDeveloperApplied to
aDry powderDry surface (dust cloud, bulb, or soft brush; remove excess gently)
bWater-solubleWet or dry surface, then dried
cWater-suspendibleWet or dry surface, then dried
dNonaqueous, for Type IDry surface, by spray (shake can well)
eNonaqueous, for Type IIDry surface, by spray
fSpecific applicationAs the manufacturer and procedure state
  • Apply a thin, even coat. A heavy coat masks fine indications.
  • Many procedures avoid water-soluble developer with Method A penetrant, because the water can draw out penetrant that already contains emulsifier.
  • Developing time has a minimum and a maximum. Watch indications as they form; it helps characterize them before they spread.

Back to lessons

8. Lighting and viewing conditions

  • Fluorescent (Type I): darkened area with very low ambient white light, UV-A of at least the required intensity at the examination surface, and dark adaptation before evaluating.
  • Visible (Type II): white light of at least the required intensity at the surface.
  • Follow the governing procedure and the lamp manufacturer's instructions for UV-A lamp warm-up.
  • Measure UV-A and white light with calibrated meters at the frequency the procedure requires.
  • Replace cracked or damaged UV-A filters. Do not wear photochromic lenses. Do not look directly into the lamp.

See the code values table for lighting intensities. Warm-up and other process times are per the governing procedure / code edition.

Back to lessons

9. Interpretation and evaluation

  • Relevant indication: caused by a discontinuity. Evaluate it against the acceptance criteria.
  • Nonrelevant indication: real cause, but from design or geometry (press fits, joints, keyways). Verify and record per procedure.
  • False indication: from processing or handling, such as contamination on gloves, rags, or tables, lint, or poor washing.
  • Defect: a discontinuity that does not meet the acceptance criteria.

Shapes: continuous lines suggest cracks, cold shuts, or lack of fusion. Intermittent lines suggest tight cracks or laps. Rounded indications suggest porosity or pits. Heavy, growing bleedout suggests more volume (often deeper).

Questionable indications: wipe lightly with a solvent-dampened swab, redevelop, and see whether it returns. Reinspection means reprocessing from pre-cleaning, not just adding developer. Excess background must be cleaned and reprocessed.

Acceptance criteria come from the governing document

  • ASME Section VIII Div. 1, Mandatory Appendix 8 defines relevant, linear, and rounded indications and lists rejectable conditions (see the code values table).
  • AWS D17.1 (aerospace fusion welding) assigns acceptance limits by weld class (A, B, or C) shown on the drawing. Evaluate against the class specified.
  • Customer specifications and drawings may add or change requirements.

Back to lessons

10. Discontinuities by product form

Product formTypical surface discontinuities
WeldsCrater cracks at stops, toe cracks, surface porosity, lack of fusion along toes, undercut
CastingsPorosity, shrinkage, cold shuts, hot tears at section changes
ForgingsLaps (folds, often intermittent), bursts that reach the surface
PlateLaminations at cut edges
Bar stock and tubeSeams running along the length
ProcessingGrinding cracks (fine network, across the grinding direction), quench cracks at corners and section changes
In serviceFatigue cracks at fillets, holes, and threads; stress corrosion cracks

Back to lessons

11. Safety

  • Read the Safety Data Sheet for each material. Aerosol penetrants, removers, and developers are often flammable: no flames or sparks, and keep cans away from heat.
  • Work with good ventilation and avoid breathing vapors and developer dust.
  • Wear chemical-resistant gloves. Penetrants and removers strip skin oils and can cause dermatitis.
  • Use UV-A lamps with intact filters, and do not look into the lamp. Wear eye protection.
  • Follow site rules for waste disposal, spill cleanup, and the halogen and sulfur limits for sensitive alloys.

Back to lessons

12. Level I and II exam topics and certification

GCT certifies personnel under its written practice, which follows SNT-TC-1A (the 2025 ASME Code references the 2024 edition) and, for aerospace work, NAS 410. Each sets its own training, experience, vision, and exam requirements; the employer's written practice controls.

GCT's PT Level II exams

  • General (closed book): principles, materials and processing, method selection, interpretation, discontinuities. The full practice mock on this site has 40 questions with a 70% target.
  • Specific (open book): the codes, specifications, and procedures you work to, such as ASME V Article 6, ASME VIII Div. 1 Mandatory Appendix 8, ASTM E1417/SE-165, and SNT-TC-1A. The practice mock has 20 questions with a 70% target.
  • Practical: hands-on processing and evaluation of real parts, done in person.

Level responsibilities

  • Level I performs specific processing and evaluations under written instructions and records results, under Level II or III guidance.
  • Level II sets up and performs the test, interprets and evaluates results against codes and specifications, reports results, and guides Level I trainees.

Vision checks (near-vision acuity and color contrast) are part of certification and are repeated at the interval the written practice sets.

Start a practice quiz

Back to lessons

Code values at a glance

Always follow the governing code, specification, and procedure, and the edition they reference. Values below are summarized in our own words for study. Where documents differ, the one that governs your work applies.

ItemASME BPVC 2025 Section V, Article 6ASTM E1417 / E1417M-21
Standard part and penetrant temperature40–125°F (5–52°C) (T-652); outside this range, qualify per T-65340–125°F (4–52°C)
Minimum penetrant dwellTable T-672: 5 min welds & castings, 10 min wrought forms (double from 40–50°F)Per the governing procedure / code edition
Maximum penetrant dwellPer the governing procedure / code editionPer the governing procedure / code edition
Developing & final interpretationPer the governing procedure / code editionPer the governing procedure / code edition
Water rinse (water-washable)Max 50 psi (350 kPa); max 110°F (43°C) (T-673.1)Per the governing procedure / code edition
Drying temperaturePer the governing procedure / code editionPer the governing procedure / code edition
UV-A at examination surfaceAt least 1000 µW/cm², checked with calibrated UV-A radiometer (T-676.4)At least 1000 µW/cm²
Ambient white light, fluorescentMax 2 fc (21.5 lx) (T-676.4)Max 2 fc (21.5 lx)
White light, visible PTAt least 100 fc (1076 lx) (T-676.3)At least 100 fc (1076 lx)
Dark adaptationAt least 5 min (T-676.4)Per the governing procedure / code edition
UV-A lamp warm-upPer the governing procedure / code editionPer the governing procedure / code edition

ASME VIII Div. 1, Mandatory Appendix 8 (acceptance)

  • Relevant: major dimension greater than 1/16 in. (1.5 mm).
  • Linear: length more than 3 times width. Rounded: length 3 times width or less.
  • Rejectable: any relevant linear indication; relevant rounded indications larger than 3/16 in. (5 mm); four or more relevant rounded indications in a line separated by 1/16 in. (1.5 mm) or less, edge to edge.
  • The size of the indication, not the imperfection, is evaluated. Reexamine questionable indications.

Train in person with GCT

Online practice helps, but nothing replaces hands-on time. GCT Inspection runs classroom and hands-on NDT training at our Pasadena, Texas facility, including Liquid Penetrant (PT) classes with hands-on practice on real parts.

Sign up for an in-person PT class or call (713) 943-1760